Fausse bisalbumienémie tricuspide : à propos d’un cas inhabituel
Type de matériel :
31
The serum protein capillary electrophoresis (SPE) is a clinical laboratory technique with high degree of resolution. However, certain interferences can be detected giving a splitting or a shoulder of the albumin fraction on the electrophoretic profile. We report an unusual finding of SPE performed with Capillarys™ (Sebia) in a 60-year-old women, hospitalized for the exploration of cholestatic jaundice, which showed an association of three anomalies in the migratory range of albumin with two supernumerary peaks, simulating bisalbuminemia. The acquired nature of bisalbuminemia was confirmed by the disappearance of all abnormalities when we treated the same sample with electrophoresis Hydragel. An interference with bilirubin, lipids and bile acids was explaining this electrophoretic aspect.
L’électrophorèse capillaire des protéines sériques (EPP) est une méthode d’analyse à très haute résolution. Toutefois, la présence de plusieurs interférents protéiques endogènes peut donner un aspect électrophorétique artéfactuel atypique de bisalbuminémie. Nous rapportons le cas atypique d’une patiente de 60 ans admise dans notre structure pour l’exploration d’un ictère choléstatique et dont l’EPP réalisée sur Capillarys™ (Sebia) révèle la présence de trois anomalies au niveau de la fraction albumine, un épaississement du bas du pic de l’albumine sur son versant anodique associé à trois pics atypiques distincts (deux pics surnuméraires du côté cathodique) ; la disparition de ces anomalies sur une seconde électrophorèse sur gel d’agarose témoigne de l’origine acquise et artéfactuelle de la bisalbuminémie.